Data Securities
The N934xC offers the optional memory erase
function for data security. Press
[System] > {More}
>
{Securities} > {Erase Memory}
to erase all the user
data in internal memory. Press Enter as a
terminator to start the erase process immediately.
Upgrading Firmware
Please follow the steps below to update the
firmware:
1
Download the latest N934xC firmware from
http://www.agilent.com/find/sa_upgrades
2
Extract files to the root directory of a USB stick.
You will see a folder named “N934xDATA” with
file
Bappupgrade.hy
.
3
Insert the USB stick into the top panel USB
connector.
4
Press
[System] > {More} > {Service} > {Upgrade
Firmware}
to activate the updating procedure.
Press
Enter
to upgrade the firmware. The
analyzer will perform the update automatically.
5
Unplug the USB stick and restart the analyzer
when message “All modules have been upgraded,
please restart” is displayed.
6
Press
[System] > {System Info} > {Show System}
to
find the updated MCU firmware version.
CAU-
CAUTION
The memory erase process takes 15 minutes
approximately. During the erase process, there must be a
constant power supply to ensure the successful erase. If
the erase process is interrupted, please reboot the
instrument and erase memory again.
CAU-
CAUTION
In updating process, there must be a constant power
supply to for at least 15 minutes. If power fails during the
updating process it can cause damage to the instrument.
Содержание N9342C
Страница 1: ... Agilent N9342C 43C 44C Handheld Spectrum Analyzer User s Guide ...
Страница 8: ......
Страница 39: ...31 AgilentN9342C 43C 44C Handheld Spectrum Analyzer 3 Functions and Measurements ...
Страница 92: ...3 Functions and Measurements 84 N9342C 43C 44C User s Guide ...
Страница 156: ...5 Error Messages 148 N9342C 43C 44C User s Guide ...
Страница 178: ...7 Menu Map 170 N9342C 43C 44C User s Guide ...