Series N6700 User’s Guide
129
Appendix B
Verification and Calibration
The verification procedures described in this appendix verify that the
Agilent N6700 Modular Power System is operating normally and is
within published specifications.
This appendix also includes calibration procedures for the power
system. Instructions are given for performing the procedures from
either the front panel or a controller over the GPIB.
NOTE
Perform the verification tests before calibrating your power system. If the power
system passes the verification tests, the unit is operating within its calibration
limits and does not need to be re-calibrated.
The recommended calibration interval for Agilent N67xxA/B power
modules is one year. Agilent N6700 MPS mainframes do not require
calibration.