3-12
Service Guide N5230-90014
Tests and Adjustments
PNA Series Microwave Network Analyzers
Preliminary Checks
N5230A
The Test Port Cable Checks
A faulty test port cable can cause a failure in the verification test. The following checks are
not required, but are recommended to verify the performance of the test port cable.
•
“Cable Return Loss Check” on page 3-13
•
“Cable Insertion Loss Check” on page 3-14
•
“Cable Magnitude and Phase Stability Check” on page 3-15
•
“Cable Connector Repeatability Check” on page 3-17
Accessories Used in the Test Port Cable Checks
Equipment Type
Model or
Part Number
Alternate Model
or Part Number
Calibration kit, 3.5 mm
85052B
85052D
Test cable, 3.5 mm (f) to 3.5 mm (f)
85131C
85131E
Содержание N5230A
Страница 10: ...Contents 6 Service Guide N5230 90014 Contents...
Страница 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Страница 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Страница 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Страница 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Страница 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Страница 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Страница 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Страница 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...
Страница 262: ...Index 14 Service Guide N5230 90014 Index...