
Service Guide N5230-90014
8-7
PNA Series Microwave Network Analyzers
General Purpose Maintenance Procedures
N5230A
Error Terms
Figure 8-4
Flowgraph of Four-Port Error Terms
where:
E = error term
DIR = Directivity
MAT = Forward Source Match and Reverse Load Match
TRK = Forward Reflection Tracking and Reverse Transmission Tracking
For the case of a full 4-port calibration, port 1 has
four Match error terms:
S11 source match
S12 load match
S13 load match
S14 load match
and four Tracking error terms:
S11 reflection tracking
S12 transmission tracking
S13 transmission tracking
S12 transmission tracking
There are eight isolation (crosstalk) terms not shown.
Содержание N5230A
Страница 10: ...Contents 6 Service Guide N5230 90014 Contents...
Страница 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Страница 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Страница 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Страница 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Страница 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Страница 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Страница 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Страница 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...
Страница 262: ...Index 14 Service Guide N5230 90014 Index...