7-32
Service Guide N5230-90014
Repair and Replacement Procedures
PNA Series Microwave Network Analyzers
Removing and Replacing the A40 Floppy Disk Drive
N5230A
Removing and Replacing the A40 Floppy Disk Drive
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16 inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
and
for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to
“Removing the Covers” on page 7-6
.
3. Remove the front panel assembly. Refer to
“Removing and Replacing the Front Panel
4. Disconnect the ribbon cable (item
①
) from the A14 system motherboard.
5. Lift the tab at the rear of the A40 floppy disk drive enclosure and remove the enclosure
rear cover (item
②
).
6. Slide the A40 floppy disk drive out of the front of the enclosure. The A40 floppy disk
drive fits snugly. You may have to push the drive from the back to remove it from the
analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure using the existing ribbon cable (item
①
).
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
Содержание N5230A
Страница 10: ...Contents 6 Service Guide N5230 90014 Contents...
Страница 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Страница 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Страница 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Страница 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Страница 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Страница 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Страница 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Страница 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...
Страница 262: ...Index 14 Service Guide N5230 90014 Index...