Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Assembly Replacement
LCD Display (N5181A/82A/83A only)
3-14
LCD Display (N5181A/82A/83A only)
Tools Required
•
T-10 driver
Removal Procedure
for this procedure.
1.
Disconnect the power cord.
2.
Remove the covers from the signal generator. Refer to
.
3.
Remove the front-panel from the signal generator. Refer to
.
4.
Remove the A6 DC-AC Inverter Interface Board from the signal generator. Refer to
”A6 DC-AC Inverter Interface Board
(N5181A/82A/83A only)” on page 3-16
5.
Using the T-10 driver, remove the six screws (1) that secure the sub-panel (2) to the front-panel.
6.
Remove the sub-panel.
7.
Remove the LCD Display.
Replacement Procedure
1.
Reverse the order of the removal procedure.
2.
Torque all screws to 9 in–lbs.
3.
Perform the post-repair performance tests that pertain to this replacement procedure, see
Chapter 4, Post-Repair Procedures and
.
深圳德标仪器