Using the Multiport Test System
How to Set Up the Multiport Test System Hardware
Chapter 3
19
5. If your network analyzer is an N8362A/B or N8364A/B PNA, you will need to perform the
Phase-Lock IF Gain Adjustment on the analyzer after it has been connected to the test set
as a system. This routine adjusts the R Channel receivers ALC gain to ensure phase lock
over the entire frequency range. Refer to
Phase-Lock IF Gain Adjustment
in the
network analyzer’s online help system for details. Use the following steps to perform this
adjustment:
a. On the network analyzer, from the
System
menu, click
Service
, then
Adjustments
,
then
IF Gain Adjustment
.
If you are unable to find these selections on your N8362A/N8364A PNA, your analyzer
firmware is a revision prior to 3.0. Follow the instructions in
to complete this adjustment.
b. Select the test set options installed.
c. No connections to the test ports are required.
d. Click
Begin Adj
. The adjustment takes about a minute to complete.
The advanced screen is for factory personnel only.
NOTE
This adjustment must be performed before using the network analyzer each
time the system is assembled or disassembled.
6. If it hasn't already been done, install and configure the software. See
Start the Balanced Measurement Software” on page 39
.
Содержание N4413A
Страница 4: ...iv ...
Страница 11: ...1 1 About the Multiport Test System ...
Страница 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Страница 19: ...9 2 User Interface ...
Страница 27: ...17 3 Using the Multiport Test System ...
Страница 89: ...79 4 Performing Time Domain Measurements ...
Страница 109: ...99 5 Menu Reference ...
Страница 163: ...153 6 Maintenance and Troubleshooting ...
Страница 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Страница 173: ...163 7 Instrument Information ...
Страница 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Страница 209: ...199 8 Safety and Regulatory Information ...
Страница 217: ...Safety and Regulatory Information Regulatory Information Chapter 8 207 Declaration of Conformity ...
Страница 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Страница 219: ...209 A Other Technical Procedures ...
Страница 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...