Menu Reference
Utility Menu
Chapter 5
150
Attenuator Tab
The
Attenuator
tab is used to manually control the attenuator that is built into the test set.
Use the
Attenuator
tab when evaluating power levels at the test-set ports and during system
checks.
Changes to the attenuator setting made here are temporary. The attenuator is reset to its
prior value when the dialog box is closed.
NOTE
Changes to the attenuator setting will degrade calibration accuracy, even after
initial settings are restored. For best results, always recalibrate after the
attenuation is changed or recycled.
Utility > Characterize Adapter…
Pull down
Utility > Characterize Adapter…
to start the adapter S-parameter
characterization procedure. After this procedure is complete, you can create a file to save the
characterization for future measurements.
See
“How to Characterize Adapters” on page 44
for more information.
Содержание N4413A
Страница 4: ...iv ...
Страница 11: ...1 1 About the Multiport Test System ...
Страница 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Страница 19: ...9 2 User Interface ...
Страница 27: ...17 3 Using the Multiport Test System ...
Страница 89: ...79 4 Performing Time Domain Measurements ...
Страница 109: ...99 5 Menu Reference ...
Страница 163: ...153 6 Maintenance and Troubleshooting ...
Страница 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Страница 173: ...163 7 Instrument Information ...
Страница 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Страница 209: ...199 8 Safety and Regulatory Information ...
Страница 217: ...Safety and Regulatory Information Regulatory Information Chapter 8 207 Declaration of Conformity ...
Страница 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Страница 219: ...209 A Other Technical Procedures ...
Страница 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...