Site Preparation
78
User Guide
Environmental Conditions
The LTM system must be operated within the recommended ranges for the gas
chromatographs.
•
Ambient operating temperature: 15°C to 35°C
•
Storage temperature extremes: –40°C to 65°C
•
Ambient operating humidity: 5% to 95% (noncondensing)
•
Altitude: Up to 5000 m.
Benchtop Space Requirements
LTM oven door
The LTM oven door replaces the existing GC oven door.
•
Height: 36.8 cm (14.5 in.)
•
Width: 43.2 cm (17.0 in.)
•
Depths: 25.4 cm (10.0 in.). With column modules installed, unit extends 18.4 cm (7.2 in.)
forward from the original door.
•
Average weight: 6.7 kg (14.7 lb)
The external power supply
•
Height: 4.6 cm (1.8 in.)
•
Width: 8.5 cm (3.3 in.)
•
Depths: 21 cm (8.3 in.). Allow 5 cm (2 in.) in front and behind for power cord connections.
•
Average weight: 1.1 kg (2.4 lb)
Some LTM systems use 2 power supplies.
Electrical Requirements
Line voltage requirements: 100–240 VAC, ± 10% of nominal