4
Operation
User Guide
63
•
For LTM Series II column modules, the GC will read primary LTM column dimensions
(length, diameter, film thickness, module/column toroid size) and temperature limits from
the LTM column module.
•
Define the column that connects the GC inlet to the LTM column module as the
In
Segment
.
•
Define the column that connects the detector to the column module as the
Out Segment
.
•
If the column segment connected to the detector passes through a second heated zone,
for example an MS transfer line, define the portion passing through the second zone as
Segment 2
.
For details on configuring columns on the GC, see the online help and
Operation Manual
available from the browser interface and in the data system.
LTM Series II columns and Agilent data systems
After installing a new column module, configure the new column module in the data system
methods.
•
Always check the data system method’s column configuration and update as needed.
•
If using a Series II column module that provides its own configuration information, the data
system will be able to edit the column information to a more limited extent. The primary
column, contained within the column module, cannot be edited for film thickness,
maximum temperatures, and certain other data. You can “calibrate” the column by slightly
changing its length or id (but do not do so unless you can accurately calculate these
dimensions).
Programming the LTM column modules
The LTM columns typically appear in the GC display as columns 1 through 4. To program the
run time column temperature and flow (or pressure) using the GC touchscreen or browser
interface go to
Method > Columns:
•
Flow, pressure, and velocity
•
Control mode (flow or pressure, ramped or not)
•
Temperature ramps
•
Post-run temperature
If using an Agilent data system, the data system method editor provides additional controls,
one for each configured LTM column module. Refer to the online help in the method editor.
LTM programming and total run time
The GC oven time controls the total run time. If an LTM column module is programmed for a
shorter or longer run time than the GC oven, the LTM column program will be extended or
truncated as needed to meet the GC oven run time.