• Every sequence can be stored in a file and recalled from that file. However, the
SequenceExpression does not include user patterns. Recalling a sequence from
its file may fail if the referenced user patterns are not available.
A recalled sequence is not automatically downloaded to the pattern generator.
• In sequence mode, the function "Error Dector Pattern tracks the Pattern
Generator Pattern" is automatically disabled. You have to download the
expected (payload) pattern to the error detector.
• During initialization of the sequence, pure zeros are sent.
User-Defined Sequences - Procedures
You can create a new sequence, save it in a file, or load a saved sequence from its
file.
Creating a New Sequence
If you wish to create a new sequence:
1
Click the
Sequence Editor.
menu item from the
Pattern
submenu.
By default, the
Sequence Editor
shows one block that is automatically repeated.
It has a length of 512 bits and generates Pause0 data (pure zeros).
2
Click the
Properties
icon.
In the
Properties
window, set:
– the sequence start condition
– the number of blocks contained in the sequence (1 to 4)
3
Click
Ok
to close the
Properties
window.
4
Specify the data to be generated during execution of a block.
– Click the
Edit
button of the block.
– Choose from the list. For example, you can select a PRBS polynomial or
browse for a pattern file. A "None" block is completely ignored.
– If necessary, change the block length. The block length of user patterns is
derived from the file. Remember that such patterns are automatically
repeated until a multiple of 512 bits is reached.
Setting up Patterns
3
Agilent J-BERT N4903B High-Performance Serial BERT
99
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...