This opens the
Self Test Options
dialog box.
2
Select the desired tests and click the
Run Test
button.
3
After the self test has finished, the results are displayed next to each selected
test.
4
If a test failed, you can view detailed error messages by clicking the respective
Message
button.
Self Test - Reference
This section contains information about the
Self Test Options
dialog box used for
setting up and running a self test and for viewing the results.
Self Test Options
The
Self Test Options
dialog box offers to select the following tests:
•
Pattern Generator Bios Self Test
This test is checking all the modules, chips, and registers of the pattern
generator more thoroughly than the Power Up test. It may take about 5 minutes
to complete.
•
Pattern Generator Power Up Messages
This test is automatically run after switching on the instrument. It mainly
checks if all important modules, chips, and registers of the pattern generator
are available and accessible.
•
Pattern Generator Auto Calibration
This test runs a timing calibration that calibrates the clock to data alignment
of the pattern generator. It may take about 45 minutes to complete.
•
Error Detector Bios Self Test
This test is checking all the modules, chips, and registers of the error
detector more thoroughly than the Power Up test. It may take about 5 minutes
to complete.
•
Error Detector Power Up Messages
This test is automatically run after switching on the instrument. It mainly
checks if all important modules, chips, and registers of the error detector are
available and accessible.
•
Error Detector Auto Calibration
This test runs a timing calibration that calibrates the clock to data alignment
of the error detector. It may take about 45 minutes to complete.
10
Customizing the Instrument
564
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...