connect the pattern generator to the N4876A. Make sure to connect Data Out
to Data In, Aux Data Out to Aux Data In and Aux Clk Out to Aux Clk In.
2
Mount the SMA 50 Ohm terminations on the unused Data Out and Aux Data
Out ports of the pattern generator using the SMA to 2.4 mm adapters.
3
Enable the N4876A as described in
Multiplexer Function” on page 47
and set the levels such that the DUT will not
be damaged.
4
Connect the differential outputs of the N4876A to the DUT. If only one output
of the N4876A is needed, connect a 2.4 mm 50 Ohm termination to the unused
output.
Installing the M8061A - Procedure
The M8061A 28 Gb/s Multiplexer with De-emphasis is an optional instrument that
can be connected to the J-BERT N4903B. For general information see
28 Gb/s Multiplexer with De-emphasis” on page 17
.
The M8061A software is integrated in the J-BERT N4903B installer, which means
that you don't require to install the M8061A software, separately. It automatically
gets installed when you install J-BERT N4903B software. The J-BERT N4903B
software supports the M8061A starting from revision 7.50.
Checking the J-BERT N4903B Software Revision
To check the J-BERT N4903B software revision:
1
Switch on the J-BERT N4903B.
2
In the
Help
menu, click
About
.
3
If the software revision is below 7.50, you need to update the software (see
“Updating the Software” on page 528
Connecting the M8061A with J-BERT N4903B via USB
The M8061A is controlled by the J-BERT N4903B via USB. Use the matched USB
cable assembly (USB 28-AWG 2.0 standard male A to B mini 4-conductor 3-mm)
to establish the USB connection. The mini USB port is available at the front side
of the AXI frame which is mounting the M8061A module. Use the USB port (USB
2.0) that is available in the rear side of the J-BERT N4903B.
Follow the given steps to establish a USB connection between M8061A and J-
BERT N4903B:
Customizing the Instrument
10
Agilent J-BERT N4903B High-Performance Serial BERT
549
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...