M8061A System Setup Adjustment
M8061A System Setup Adjustment
The input timing of M8061A has to be adjusted to ensure error free operation for
all data rates. This adjustment is required once for a dedicated setup of N4903B,
M8061A and the cable kit being used, and has to be repeated whenever one of
these are changed in the setup or the operating temperature is changed by more
than 5
o
Celsius, compared to the previous adjustment.
Input timing alignment can be started from "Adjustment" button. The type of
alignment depends on the clock source selected.
1
When the clock source is an external clock (clean clock mode), using the “CLK
IN” connector on the front panel, realignment needs to take place each time
the frequency, cable set or setup is changed. Results are not saved.
2
When clock source is "AUX CLK IN" internal clock, the results are saved once
the alignment is done.
N O T E
Ensure that the normal and complement data outputs of the M8061A are terminated
with 50 Ohm before starting the alignment.
Performing M8061A System Setup Adjustment Setting
Follow the given steps to perform system setup adjustment setting:
1
From the
Navigation Menu
, select
External Instrument(s)
and then click
Config
sub-menu. The
Config
window shows a list of all instruments connected
to the N4903B.
2
Select the external instrument from the given list. For example, here you will
select
M8061A
.
3
Click
Adjustment...
to continue the system setup adjustment setting.
A message will pop up which warns you to save the instrument settings before
continuing the auto adjustment process.
4
Click
Yes
to continue if you have already saved the settings. Once the auto
adjustment starts, you will see the adjustment progress in the following screen:
9
Solving Problems
492
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...