N4916B System Setup Calibration
N4916B System Setup Calibration
The input timing of the N4916B has to be calibrated to ensure error free operation
for all data rates. This calibration is required once for a dedicated setup of N4903B,
N4916B and the cable kit being used, and has to be repeated whenever one of
these are changed in the setup.
N O T E
The Input Timing setting that is determined during this calibration is only valid for
all data rates when using the N4915A-010 cable kit.
For all other cables, this calibration has to be done for each data rate that has to be
set using the N4916B De-Emphasis Signal Converter.
Maintaining a list of proper Input Timing Settings to data rates and use of the correct
setting for a dedicated data rate is up to the user when not using the cable kit.
Loading Calibration Settings
Follow the given steps to load the calibration setting:
1
From the
Navigation Menu
, select
External Instrument(s)
and then click
Config
sub-menu. The
Config
window shows a list of all instruments connected
to the N4903B.
2
Select the external instrument from the given list. For example, here you will
select
N4916B
.
3
Use the drop down list to provide the
Input Timing Setup
.
4
Click
Load Calibration Settings
to continue. A message will pop up which
warns you to save the instrument settings before continuing the calibration.
9
Solving Problems
480
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...