If you have enabled sinusoidal interference, enter appropriate values for
Frequency
,
Amplitude
, and
Mode
.
Interference Channel Setup - Reference
The Interference Channel module must be enabled before it can be used. This is
done by clicking the corresponding button.
When the Interference Channel is disabled, it establishes an almost direct
connection between P1 and P2. For details about the minimum trace length refer
to the technical specifications.
Trace Number
To specify the degree of intersymbol interference, select a suitable trace length.
The user interface provides numbers from 1 to 9. The greater the number, the longer
is the trace.
Sinusoidal Interference
To enable the addition of sinusoidal interference, press the corresponding button.
Set the parameters to desired values:
The peak-to-peak amplitude of sinusoidal interference must not exceed a certain
level (see the technical specifications).
8
Jitter Tolerance Tests
432
Agilent J-BERT N4903B High-Performance Serial BERT
Amplitude
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...