Errored 0's Ratio
The ratio of the errored 0's count to the bit count is displayed here to 4
significant digits.
•
Errored 0's Count
The number of logic 0's measured as logic 1's during the accumulation period
is displayed here.
•
Auto Resync Counter
The total number of Auto Re-Sync Count received in a time interval is displayed
here.
G.821 Measurements
•
Availability
The ratio of the available seconds to the total gating period expressed as a
percentage.
•
Unavailability
The error ratio is calculated over 1 second timed intervals during the gating
period. An unavailable period begins when the error ratio is worse than 1 x
10
-3
for 10 consecutive seconds. These 10 seconds are considered part of the
unavailable time. The unavailable period ends when the error ratio is better
than 1 x 10
-3
for 10 consecutive seconds. These 10 seconds are considered part
of the available time. Percent (%) Unavailability is the ratio of the unavailable
seconds to the total gating period expressed as a percentage.
•
Severely Errored Seconds
The ratio of the total number of available seconds with an error ratio worse
than 1 x 10
-3
to the total number of available seconds, expressed as a
percentage.
•
Degraded Minutes
Severely errored seconds are discarded from the available time and the
remaining seconds are grouped into blocks of 60 seconds. Blocks which have
an error ratio worse than 1 x 10
-6
are called Degraded Minutes and Percent (%)
Degraded Minutes is the ratio of the total number of degraded minutes to the
total number of 60 second blocks in the available time expressed as a
percentage. Incomplete blocks of less than 60 seconds are treated as complete
blocks of 60 seconds.
•
Errored Seconds
The ratio of the errored seconds in the available time to the total number of
seconds in the available time, expressed as a percentage.
Evaluating Results
7
Agilent J-BERT N4903B High-Performance Serial BERT
381
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...