3
View the
Bit Comparison without USB3.1 SKPOS
bar and errors indicator. See
“Status Indicators” on page 372
How to Analyze Instantaneous Measurements Results
To analyze the behavior of your device, you can do the following modifications to
the test setup and monitor the effect on the resulting "BER" / "8B/10B
Comparison" / "Bit Comparison without PCIe3 SKPOS" / "Bit Comparison without
USB3.1 SKPOS":
1
Add errors to the data stream. See
“Manually Inserting Errors” on page 157
details.
2
Use alternating patterns to switch between different data streams. See
Up a Periodical Output” on page 162
3
Adjust the parameters of the output port of the pattern generator, as described
“Setting Logic Levels and Terminations” on page 131
4
Manually move the sampling point. See
“How to Set the Optimum Sampling
Point Automatically ” on page 196
for details.
5
Make adjustments to your device.
This can be used to see how changes to your device affect the "BER" / "8B/
10B Comparison" / "Bit Comparison without PCIe3 SKPOS" / "Bit Comparison
without USB3.1 SKPOS". You can determine what adjustments improve or
degrade the "BER" / "8B/10B Comparison" / "Bit Comparison without PCIe3
SKPOS" / "Bit Comparison without USB3.1 SKPOS".
Evaluating Results
7
Agilent J-BERT N4903B High-Performance Serial BERT
365
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...