The ‘Error’ option gives the user the choice of saving, the BER, the Compared Bits,
and the Errors, for different combinations of expected 1s, 0s, and all.
Eye Diagram - Reference
The Eye Diagram measurement returns the results in graphical and numerical
forms. The following section explains the measured parameters, and the displayed
options that are specific to these measurements. Additionally, some information
is provided to explain the theoretical background to these measurements.
Parameters that can be set on the various tabs of the Properties pop-up:
Parameters
N O T E
If the parameters on this page are modified, run the measurement again to update
the results. While the measurement is running the parameter settings cannot be
changed.
Set the criteria for Persistence.
• Infinite
With ‘Infinite’, the eye appears with the Automated Eye Parameter Measurement,
and the results keep on improving in accuracy, as more number of points get
measured and represented on the diagram along with the deeper BER values. The
BER threshold in the Graph Tab decides how long the measurements will be done
at the boundaries of the eye. Once the measured points exceed the BER Threshold
defined by the user, the measurement is done deep inside the eye.
This is a default setup for persistence.
• Minimum
With ‘minimum’, a reasonably clear eye, with transition, appears along with all the
Automated Eye Parameter Measurement, and automatically the measurement
restarts showing only the refined eye diagram. To stop the measurements click the
abort button.
• Secs
With ‘Secs’, the measurement runs for the specified time, and the measurement
automatically restarts showing only the refined eye diagram. To stop the
measurement, click the abort button.
Sets the transition of the eye on the plot
• Middle of eye
6
Advanced Analysis
332
Agilent J-BERT N4903B High-Performance Serial BERT
Persistence
Center to
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...