Connecting the DUT
Connecting the DUT - Concepts
This section provides information on how to connect your DUT for several common
test scenarios.
N O T E
The pattern generator's Data Out, Aux Data Out, Clock Out and Trigger/Ref Clock
Out ports must be terminated with 50
Ω
if they are not connected.
N O T E
The error detector's Clock In connector must be terminated with 50
Ω
if they are
not in use.
The following procedure is recommended when setting up a test:
1
If your DUT can handle 0V, press the
0V (Disable)
button in the
PG Setup
-
Data Output
or
PG Setup
-
Clock/Trigger Output
screen.
The pattern generator's Data Out, Aux Data Out, Clock Out and Trigger/Ref
Clock Out outputs are set to 0V. This is indicated in the
PG Setup
-
Data
Output
or
PG Setup
-
Clock/Trigger Output
screen.
2
Connect the DUT as necessary.
3
Terminate any non-connected Data Out, Aux Data Out, Clock Out and Trigger/
Ref Clock Out ports (normal and complementary).
4
If the outputs are disabled, press the
0V (Disable)
button to enable them.
To avoid damaging your device, set up the devices properly before making
connections.
C A U T I O N
ESD can damage or destroy electronic components. Coaxial cables with both ends
unconnected may store electrostatic charges. Before connecting any coaxial cable
of this sort to a device or instrument, momentarily short the center and outer
conductors of the cable. When making connections, ensure the proper use of a
grounded, resistor-isolated wrist strap.
Connect your device according to the diagrams below:
Planning the Test
1
Agilent J-BERT N4903B High-Performance Serial BERT
29
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...