This graph shows the extrapolation of the optimum Q-factor and the optimum
threshold level from a limited number of measured points.
The measured data points to be used for the calculation have to be within a
contiguous BER range. This range is defined by specifying the
Min BER for Q
(lower
threshold) and the
BER Threshold
(upper threshold). Both thresholds can be set
in the lower section of the
View
tab.
The Q-factor can only be calculated if for both high and low level rails two or more
points fall within the defined BER range. For reliable results use at least five
measured points.
To ensure proper settings, activate the
Show Measured Points
function on the
Graph
tab and switch to the
BER vs. Threshold
graph. Move the upper BER
threshold marker (vertical line) so that a sufficient number of measured points is
included for the calculation:
From these points, the following
QBER vs. Threshold
graph is generated:
This graph illustrates the calculation of the
Q-factor
(65 in the figure above) as the
best fit line through the calculated points. This method makes the calculation
reproducible. For details see
“Q-factor Results” on page 279
Furthermore, the following parameters can be set on the
View
tab of the
Properties
dialog box:
You can calculate measurement parameters for:
•
BER Threshold
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
275
QBER vs. Threshold Graph
Calculate
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...