But your dBER distribution may also look like this:
The measurement will calculate the Level and Standard Deviation results from all
data points. The marker allows you to measure the μ (Mu) and σ (Sigma) of the
individual peaks.
DUT Output Timing/Jitter
DUT Output Timing/Jitter - Concepts
This type of measurement is used to measure the timing and jitter behavior for a
device under test (DUT). It uses a bit error rate (BER) measurement to evaluate the
shape of the eye for the output signal of the DUT. It also analyzes the jitter,
separates the random jitter and deterministic jitter components, and estimates the
total jitter.
A direct result is the determination of the optimum sampling point delay for
receiving data from the DUT with maximum confidence.
DUT Output Timing/Jitter includes the Fast Total Jitter measurement that can be
used to measure the total jitter for devices which generate a very low error density
in a reasonable time span.
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
241
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...