Why Can Wrong Terminations Damage Your Device?
Choosing wrong terminations may cause your device to output voltage levels that
are not as expected. It may also cause excessive current or current flow in the
wrong direction, which can damage your device.
N O T E
The clock termination is set to 50 Ohm, AC-coupled, and cannot be changed.
Inputs and Outputs - Procedures
To select the termination for the error detector:
N O T E
Prerequisite: You must know the termination voltage of the data signal that your
DUT sends to the error detector.
1
In the
ED Setup
menu, press
Sampling Point Setup
to access the
Sampling
Point Setup
window.
2
Click the
Edit
button. This opens the
ED Input Setup
dialog.
3
In the
Termination
field, enter the termination voltage that is appropriate for
the incoming data signal.
You can use the front panel knob or the numeric keypad to enter the value.
C A U T I O N
Selecting the wrong termination may damage your device.
4
Select the
Data Inverted
checkbox if your device inverts data.
You can now physically connect the DUT to the error detector.
Setting up the Error Detector
5
Agilent J-BERT N4903B High-Performance Serial BERT
171
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...