Internal
Select this option to use internal error insertion functions:
•
Error Rate
Allows to select from a preset bit error ratio from
1.000E-02
to
1.000E-09
. Errors
will be added to the output pattern to produce this error ratio.
•
Average Number of Bits Between Errors
Allows you to enter the average number of valid bits between two errors. This
is the inverse of the bit error ratio.
Off
Select this option turn off all internal add rates and external error inputs. This allows
you to manually add single errors by pressing
Error Add
on the main display. This
flips a single bit in the output stream.
Pattern Alternation
Pattern Alternation - Concepts
This section explains the basic purpose of alternating patterns. Two alternate
patterns can be stored in one pattern file.
Note that you can also generate a sequence of up to four alternating patterns (see
“User-Defined Sequences - Concepts ” on page 96
).
Standard and Alternating Patterns
With the Serial BERT, you have full control on the data stream that is sent by the
pattern generator to your device under test. You can use standard patterns to
specify a test data pattern that is sent to your DUT repeatedly.
Or, you can use alternating patterns to specify two patterns that are sent to the
DUT alternately. To implement alternating patterns, the instrument's user pattern
memory is split into two portions; one for pattern A and one for pattern B.
The differences between these patterns are:
• Standard patterns consist of one single pattern that is repeatedly sent. Any
arbitrary pattern up to 32Mbit length can be set up. However, there are no
Setting up the Pattern Generator
4
Agilent J-BERT N4903B High-Performance Serial BERT
161
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...