Patterns: P1 (Pause1), undistorted PRBS, a user pattern from a file,
or a fraction of the clock pulse (divider range: 2 ... 127).
•
Length: Must be 512 bits or a multiple thereof. Shorter or longer user
patterns are automatically repeated until the next 512-bit boundary
is reached (see also
“How the Serial BERT Generates Memory-Based
•
Trigger: Refers to the pattern generator's Trigger Out port. If the
Trigger Out is set up to generate a divided clock pulse, the block
setting is ignored.
But if the Trigger Out is set to "Sequence Trigger" and the block to
TrigOn, a trigger pulse is generated whenever the block starts or
restarts. No trigger is sent if the block is set to TrigOff.
“Trigger Output - Procedures” on page 94
for information on
how to set the mode of the Trigger Out port.
Default is one overall endless loop for the whole sequence.
•
B
n
, B
m
: Define the start and end block of a user-defined loop.
Loops always define the transition from the end of a block to the
beginning of the same or a previous block. It is not possible to jump
into an existing loop. It is also not possible to specify loops within
loops (exept the default overall loop).
•
Loop end condition: Default is an INFinite (endless) loop. Other
options are:
– AuxInHi | AuxInLo: Loop continues until Auxiliary Input is high
or low.
– AuxInRising | AuxInFalling: Loop continues until Auxiliary
Input receives a rising or falling edge.
– Manual: Loop continues until the
Break
button is clicked.
– Counted loop: Loop repeats a specified number of times.
Sequence Mode Characteristics
When a sequence is executed, the pattern generator is in "Sequence
Mode".
This mode differs from usual "Pattern Mode":
•
Pattern-related triggers cannot be generated. In sequence mode, the
Trigger Out port can produce either a continuous pulse (a fraction
of the current clock pulse) or a trigger spike at the beginning of each
block that is set to TrigOn. The duration of that spike depends on
several conditions and is therefore undefined.
2
Setting up Patterns
58
Agilent J-BERT N4903 High-Performance Serial BERT
LOOP Options
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...