![Agilent Technologies J-BERT N4903 Скачать руководство пользователя страница 427](http://html1.mh-extra.com/html/agilent-technologies/j-bert-n4903/j-bert-n4903_user-manual_2868694427.webp)
4
If a test failed, you can view detailed error messages by clicking the
respective
Message
button.
Self Test - Reference
This section contains information about the
Self Test Options
dialog box
used for setting up and running a self test and for viewing the results.
Self Test Options
The
Self Test Options
dialog box offers to select the following tests:
•
Pattern Generator Bios Self Test
This test is checking all the modules, chips, and registers of the
pattern generator
more thoroughly than the Power Up test. It may
take about 5 minutes to complete.
•
Pattern Generator Power Up Messages
This test is automatically run after switching on the instrument. It
mainly checks if all important modules, chips, and registers of the
pattern generator
are available and accessible.
•
Pattern Generator Auto Calibration
This test runs a timing calibration that calibrates the clock to data
alignment of the
pattern generator
. It may take about 15 minutes to
complete.
•
Error Detector Bios Self Test
This test is checking all the modules, chips, and registers of the
error
detector
more thoroughly than the Power Up test. It may take about
5 minutes to complete.
•
Error Detector Power Up Messages
This test is automatically run after switching on the instrument. It
mainly checks if all important modules, chips, and registers of the
error detector
are available and accessible.
•
Error Detector Auto Calibration
This test runs a timing calibration that calibrates the clock to data
alignment of the
error detector
. It may take about 15 minutes to
complete.
Customizing the Instrument
9
Agilent J-BERT N4903 High-Performance Serial BERT
427
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...