42
Agilent Intuvo 9000 GC Installation
1
Installing the GC
Prepare gas supplies.
Most of installation involves plumbing gas to tanks, filters, and
flow modules. Swagelok fittings are used to make leak-tight
connections. If you are not sure how to make a Swagelok
connection, see
for instructions.
Install the gas regulators
1
Select the appropriate CGA regulator for each gas type. (In
other countries, refer to local standards. See the Agilent
Intuvo 9000 GC, GC/MS, and ALS Site Preparation Guide
for
requirements.)
2
Confirm that the outlet fitting of the regulator is 1/8-inch
Swagelok. If not, install the appropriate adapter fitting. Wrap
the fitting threads using PTFE tape. Wrap the tape in a
clockwise direction so that the adapter threads do not
unwrap the tape. Be careful to keep the tape away from the
end of the fitting. Two to three tightly wound wraps are
sufficient.
Never use a liquid thread sealant.
Liquid thread
sealants introduce contamination into the GC plumbing
system. Tighten the Swagelok adapter fitting securely to the
NPT pipe thread fitting.
WARNING
Hydrogen is a flammable gas. If hydrogen or any other flammable
gas is used, periodic leak tests should be performed. Be sure that
the hydrogen supply is off until all connections are made.
Substituting parts or performing any unauthorized modification to
the instrument may result in a safety hazard.
Table 5
Gas regulators, 1/8-inch, U.S. only
*
* For 1/4-inch tubing, purchase a 1/4-inch to 1/8-inch adapter, U.S. only.
Description
Part number
CGA 346, 125 psig max (8.6 bar), Air
5183-4641
CGA 350, 125 psig max (8.6 bar), H2, Ar/Me
5183-4642
CGA 540, 125 psig max (8.6 bar), O2
5183-4643
CGA 580, 125 psig max (8.6 bar), He, Ar, N2
5183-4644
CGA 590, 125 psig max (8.6 bar), Air
5183-4645
Содержание Intuvo 9000
Страница 1: ...Agilent Technologies Agilent Intuvo 9000 Gas Chromatograph Installation and First Startup...
Страница 6: ...6 Agilent Intuvo 9000 GC Installation...
Страница 25: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 25 3 Reinstall the GC covers...
Страница 26: ...26 Agilent Intuvo 9000 GC Installation 1 Installing the GC Connect the power cord and LAN cable LAN...
Страница 27: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 27 Turn on the GC...
Страница 29: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 29...
Страница 52: ...52 Agilent Intuvo 9000 GC Installation 1 Installing the GC Detector cover Vent chimney Screw O Ring...
Страница 84: ...84 Agilent Intuvo 9000 GC Installation 1 Installing the GC 8 Close the GC front door...
Страница 114: ...Agilent Technologies...