Agilent InfinityLab ELSD User Manual
77
7
Error Information
Module Specific Error Messages
Rear Vapor sensor failed
Error ID: 13
The vapor sensor located on the rear panel of the module has failed.
Vapor detected
Error ID: 14
Leak detected
Error ID: 15
A leak was detected inside the module.
Probable cause
Suggested actions
1
Vapor sensor not connected to the
mainboard.
Please contact your Agilent service
representative.
2
Defective vapor sensor.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
External vapor being drawn into unit.
Remove any source of solvent vapors close to
the module.
2
Solvent leak inside unit.
Please contact your Agilent service
representative.
3
Exhaust tube not fitted.
Fit black exhaust tube.
Probable cause
Suggested actions
1
Loose nebulizer fittings.
Please contact your Agilent service
representative.
2
Blocked nebulizer causing leak at capillary
fittings.
Please contact your Agilent service
representative.