4
Using the Module
Instrument Controls
Agilent InfinityLab ELSD User Manual
46
Status Mode
The ELS Detector can be operated in two modes;
STANDBY
or
RUN
, both of which
are described over page:
To display the current mode and/or select a new mode, highlight the
MODE
function on the instrument display. The current mode will now be displayed on
the screen. Using the arrow keys, scroll up or down until the desired option is
displayed. The instrument acknowledges the command by displaying the mode
of operation in the top-right corner of the screen.
Standby
The
STANDBY
mode is the “ground state” of the ELS detector, which is by default
selected automatically after power-on (default can be changed using Power
Mode,
STANDBY
mode the heaters and light
source are switched off, and the gas manifold valve is closed at power-on. The
STANDBY
mode gives the user a control platform in which to set up the
operational parameters (gas flow, nebulizer, and evaporator temperatures)
before switching the unit into
RUN
mode. The instrument will default to
STANDBY
mode should an error occur on the instrument.
Following a command or error, the instrument is switched from
RUN
to
STANDBY
mode. In this situation, the gas management system is invoked and the gas flow
set to a minimum flow of 1.2 SLM for 15 min before the gas manifold valve is
closed. This minimum “blanket” gas is
STANDBY
enough to nebulize and
evacuate solvent should the instrument default to
STANDBY
mode with solvent
still flowing.
RUN
The
RUN
mode is the detector’s operational mode. In this mode, the instrument is
controlled at the set temperatures and gas flow, and the system is fully
operational. During heating or cooling the instrument will display
NOT READY
to
show that the system has not reached the set conditions. When the instrument
has equilibrated,
READY
will be displayed and the instrument is ready for use.
CAUTION
Flooding the detector
If the instrument is left in STANDBY mode for longer than 15 min, gas flow to
the unit is stopped to minimize gas usage.
The solvent pump must be turned off if the ELSD is going to be left in
STANDBY mode longer than 15 min to prevent solvent flooding the detector.