98
Agilent 1290 Infinity Binary Pump User Manual
7
Error Information
Pump Error Messages
Pump drive stroke blocked
Error ID: 22441
The pump drive movement is blocked.
•
Parameter: 1 – 4 referring to pump drive
Pump drive stop not found
Error ID: 22442
The pump drive stop has not been found.
•
Parameter: 1 – 4 referring to pump drive
Pressure sensor calibration wrong or missing
Error ID: 22443
Pressure sensor calibration wrong or missing.
•
Parameter: None
Probable cause
Suggested actions
1
Pump drive motor is mechanically blocked.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Pump drive spindle is defect.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Pressure sensor calibration wrong or
missing.
•
Replace pressure sensor.
•
Please contact your Agilent service
representative.
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...