Error Information
Pump Error Messages
Pressure below lower limit
Error ID: 22015
The pressure has dropped below the lower limit.
•
Parameter: None
Target pressure not reached for binary pump degasser
Error ID: 22031
The target pressure of the binary pump degasser has not been reached within
the expected time.
•
Parameter: Pressure in mbar
Probable cause
Suggested actions
1
Leak
Check for leaks.
2
Bottle empty
Check bottle filling.
3
Wrong solvent (viscosity)
Check solvent.
4
Inappropriate setting
Check flow rate and lower pressure limit.
5
Column degradation
Replace column.
Probable cause
Suggested actions
1
Condensation in degasser chamber due to
temperature fluctuation.
Equilibrate and restart module.
2
Degasser is defect.
Please contact your Agilent service
representative.
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...