Agilent 1290 Infinity Binary Pump User Manual
9
Maintenance
Replacing the Pressure Sensor
Replacing the Pressure Sensor
When
No or invalid pressure signal
Tools required
p/n
Description
8710-2412
Hex key 2.5 mm, 15 cm long, straight handle
8710-0510
Wrench open 1/4 — 5/16 inch
Screwdriver
Parts required
#
p/n
Description
1
G4220-60001
Pressure sensor 1200 bar
Preparations
Turn off pump flow, switch off pump
N O T E
This procedure describes how to replace the pressure sensor.
In case the cable to the sensor shall be replaced as well, please contact your Agilent
service representative.
1
Remove capillary connections between the pressure
sensor and purge valve.
2
Remove the screws that fix the pressure sensor to the
chassis.
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...