
Front-Panel Controls
2
6000 Series Oscilloscope User’s Guide
79
Example
Connect the oscilloscope probes for channels 1 and 2 to the
Probe Comp output on the front panel of the instrument. Be
sure to connect the probe ground leads to the ground lug beside
the Probe Comp output. Set the instrument to the factory
default configuration by pressing the
Save/Recall
key, then the
Default Setup
softkey. Then press the
AutoScale
key. You should
see a display similar to the one shown below.
Figure 9
AutoScale of oscilloscope channels 1 and 2
For an explanation of how AutoScale works, see
To set the probe attenuation factor
Passive Probes
All 6000 Series oscilloscopes including the 100 MHz models
recognize passive probes such as the 10073C and 10074C. These
probes have a pin on their connector that connects to the ring
around the oscilloscope’s BNC connector. Therefore, the
oscilloscope will automatically set the attenuation factor for
recognized Agilent passive probes.
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......