244
6000 Series Oscilloscope User’s Guide
5
Making Measurements
Amplitude
The Amplitude of a waveform is the difference between its
Top and Base values. The Y cursors show the values being
measured.
Average
Average is the sum of the waveform samples divided by the
number of samples over one or more full periods. If less
than one period is displayed, Average is calculated on the
full width of the display. The X cursors show what portion
of the displayed waveform is being measured.
Base
The Base of a waveform is the mode (most common value)
of the lower part of the waveform, or if the mode is not well
defined, the base is the same as Minimum. The Y cursor
shows the value being measured.
Maximum
Maximum is the highest value in the waveform display. The
Y cursor shows the value being measured.
Minimum
Minimum is the lowest value in the waveform display. The Y
cursor shows the value being measured.
Digital channel voltage measurements
Automatic voltage measurements are not valid on digital channels on the
mixed-signal oscilloscope.
Average =
Σ
x
i
n
---------
where
x
i
= value at
i
th point being measured
n
= number of points in measurement interval
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......