190
6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
To trigger on all sync pulses
To quickly find maximum video levels, you could trigger on all
sync pulses. When
All Lines
is selected as the TV trigger mode,
the oscilloscope will trigger on all horizontal sync pulses.
1
Press the Trigger
More
key, then press the
TV
softkey.
2
Press the
Settings
softkey, then press the
Standard
softkey to
select the appropriate TV standard.
3
Press the
Mode
softkey and select
All Lines
.
Figure 16
Triggering on All Lines