102
6000 Series Oscilloscope User’s Guide
2
Front-Panel Controls
To perform service functions
•
Press the
Utility
&
Service
to display the service menu.
The Service Menu lets you:
•
Perform User Cal on the oscilloscope.
•
View User Cal status.
•
Perform instrument Self Test.
•
View information about your oscilloscope model number,
code revision information, and User Cal status.
User Calibration
Perform user-calibration:
•
Each year or after 2000 hours of operation.
•
If the ambient temperature is >10° C from the calibration
temperature.
•
If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience
with other instruments help determine if you need shorter User
Cal intervals.
User Cal performs an internal self-alignment routine to
optimize the signal path in the oscilloscope. The routine uses
internally generated signals to optimize circuits that affect
channel sensitivity, offset, and trigger parameters. Disconnect
all inputs and allow the oscilloscope to warm up before
performing this procedure.
Start
User Cal
Start
Self Test
Return to
previous menu
About
Oscilloscope
User Cal
status
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......