
Obsolete and Discontinued Commands
7
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
703
:EXTernal:PMODe
Command Syntax
:EXTernal:PMODe <pmode value>
<pmode value> ::= {AUTo | MANual}
The probe sense mode is controlled internally and cannot be set. If a
probe with sense is connected to the specified channel, auto sensing is
enabled; otherwise, the mode is manual.
If the pmode sent matches the oscilloscope's setting, the command will be
accepted. Otherwise, a setting conflict error is generated.
Query Syntax
:EXTernal:PMODe?
The :EXTernal:PMODe? query returns AUT if an autosense probe is
attached and MAN otherwise.
Return Format
<pmode value><NL>
<pmode value> ::= {AUT | MAN}
N O T E
The :EXTernal:PMODe command is an obsolete command provided for compatibility to
previous oscilloscopes.
Содержание InfiniiVision 7000A Series
Страница 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Страница 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 652: ...652 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 5 Commands by Subsystem...
Страница 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Страница 784: ...784 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 10 Synchronizing Acquisitions...
Страница 810: ...810 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 11 More About Oscilloscope Commands...
Страница 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...