:WAVeform Commands
31
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
945
:WAVeform:SOURce
Command Syntax
:WAVeform:SOURce <source>
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r> | SBUS{1 | 2}}
for DSO models
<source> ::= {CHANnel<n> | POD{1 | 2} | BUS{1 | 2} | FUNCtion
| MATH | WMEMory<r> | SBUS{1 | 2}}
for MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= {1 | 2}
The :WAVeform:SOURce command selects the analog channel, function,
digital pod, digital bus, reference waveform, or serial decode bus to be
used as the source for the :WAVeform commands.
Function capabilities include add, subtract, multiply, integrate,
differentiate, and FFT (Fast Fourier Transform) operations.
When the :WAVeform:SOURce is the serial decode bus (SBUS1 or SBUS2),
ASCii is the only waveform format allowed, and the :WAVeform:DATA?
query returns a string with timestamps and associated bus decode
information.
With MSO oscilloscope models, you can choose a POD or BUS as the
waveform source. There are some differences between POD and BUS when
formatting and getting data from the oscilloscope:
•
When POD1 or POD2 is selected as the waveform source, you can
choose the BYTE, WORD, or ASCii formats (see
When the WORD format is chosen, every other data byte will be 0. The
setting of :WAVeform:BYTeorder controls which byte is 0.
When the ASCii format is chosen, the :WAVeform:DATA? query returns a
string with unsigned decimal values separated by commas.
•
When BUS1 or BUS2 is selected as the waveform source, you can
choose the WORD or ASCii formats (but not BYTE because bus values
are always returned as 16- bit values).
When the ASCii format is chosen, the :WAVeform:DATA? query returns a
string with hexadecimal bus values, for example: 0x1938,0xff38,...
Query Syntax
:WAVeform:SOURce?
The :WAVeform:SOURce? query returns the currently selected source for
the WAVeform commands.
Содержание InfiniiVision 3000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 152: ...152 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Страница 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 306: ...306 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 14 EXTernal Trigger Commands...
Страница 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Страница 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Страница 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Страница 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Страница 466: ...466 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 20 MEASure Power Commands...
Страница 502: ...502 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 22 POD Commands POD1 DISP 0 THR 1 40E 00...
Страница 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Страница 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Страница 1046: ...1046 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 34 Obsolete and Discontinued Commands...
Страница 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Страница 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Страница 1086: ...1086 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 37 Synchronizing Acquisitions...
Страница 1216: ...1216 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 39 Programming Examples...
Страница 1231: ...Index Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 1231 zoomed window horizontal scale 841...
Страница 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...