62
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
3
Getting Started
When you send the :DIGitize command to the oscilloscope, the specified
channel signal is digitized with the current :ACQuire parameters. To obtain
waveform data, you must specify the :WAVeform parameters for the
SOURce channel, the FORMat type, and the number of POINts prior to
sending the :WAVeform:DATA? query.
The number of data points comprising a waveform varies according to the
number requested in the :ACQuire subsystem. The :ACQuire subsystem
determines the number of data points, type of acquisition, and number of
averages used by the :DIGitize command. This allows you to specify exactly
what the digitized information contains.
The following program example shows a typical setup:
myScope.WriteString ":ACQuire:TYPE AVERage"
myScope.WriteString ":ACQuire:COMPlete 100"
myScope.WriteString ":ACQuire:COUNt 8"
myScope.WriteString ":DIGitize CHANnel1"
myScope.WriteString ":WAVeform:SOURce CHANnel1"
myScope.WriteString ":WAVeform:FORMat BYTE"
myScope.WriteString ":WAVeform:POINts 500"
myScope.WriteString ":WAVeform:DATA?"
This setup places the instrument into the averaged mode with eight
averages. This means that when the :DIGitize command is received, the
command will execute until the signal has been averaged at least eight
times.
After receiving the :WAVeform:DATA? query, the instrument will start
passing the waveform information.
Digitized waveforms are passed from the instrument to the controller by
sending a numerical representation of each digitized point. The format of
the numerical representation is controlled with the :WAVeform:FORMat
command and may be selected as BYTE, WORD, or ASCii.
N O T E
Ensure New Data is Collected
When you change the oscilloscope configuration, the waveform buffers are cleared. Before
doing a measurement, send the :DIGitize command to the oscilloscope to ensure new data
has been collected.
N O T E
Set :TIMebase:MODE to MAIN when using :DIGitize
:TIMebase:MODE must be set to MAIN to perform a :DIGitize command or to perform any
:WAVeform subsystem query. A "Settings conflict" error message will be returned if these
commands are executed when MODE is set to ROLL, XY, or WINDow (zoomed). Sending the
*RST (reset) command will also set the time base mode to normal.
Содержание InfiniiVision 3000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 152: ...152 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Страница 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 306: ...306 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 14 EXTernal Trigger Commands...
Страница 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Страница 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Страница 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Страница 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Страница 466: ...466 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 20 MEASure Power Commands...
Страница 502: ...502 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 22 POD Commands POD1 DISP 0 THR 1 40E 00...
Страница 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Страница 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Страница 1046: ...1046 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 34 Obsolete and Discontinued Commands...
Страница 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Страница 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Страница 1086: ...1086 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 37 Synchronizing Acquisitions...
Страница 1216: ...1216 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 39 Programming Examples...
Страница 1231: ...Index Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 1231 zoomed window horizontal scale 841...
Страница 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...