:MEASure Commands
19
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
397
This command defines the behavior of the :MEASure:DELay? query by
specifying the start and stop edge to be used. <edge_spec1> specifies the
slope and edge number on source1. <edge_spec2> specifies the slope and
edge number on source2. The measurement is taken as:
delay = t(<edge_spec2>) - t(<edge_spec1>)
:MEASure:DEFine
THResholds
Command Syntax
:MEASure:DEFine THResholds,<threshold spec>
<threshold spec> ::= {STANdard}
| {<threshold mode>,<upper>,<middle>,<lower>}
<threshold mode> ::= {PERCent | ABSolute}
for <threshold mode> = PERCent:
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
and lower threshold percentage values
between Vbase and Vtop in NR3 format.
for <threshold mode> = ABSolute:
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
and lower threshold absolute values in
NR3 format.
•
STANdard threshold specification sets the lower, middle, and upper
measurement thresholds to 10%, 50%, and 90% values between Vbase
and Vtop.
•
Threshold mode PERCent sets the measurement thresholds to any
user- defined percentages between 5% and 95% of values between Vbase
and Vtop.
•
Threshold mode ABSolute sets the measurement thresholds to absolute
values. ABSolute thresholds are dependent on channel scaling
(:CHANnel<n>:RANGe or
page 270:CHANnel<n>:SCALe), probe attenuation (:CHANnel<n>:PROBe),
and probe units (:CHANnel<n>:UNITs). Always set these values first
before setting ABSolute thresholds.
Query Syntax
:MEASure:DEFine? <meas_spec>
<meas_spec> ::= {DELay | THResholds}
The :MEASure:DEFine? query returns the current edge specification for the
delay measurements setup or the current specification for the thresholds
setup.
N O T E
The :MEASure:DELay command and the front-panel delay measurement use an auto-edge
selection method to determine the actual edge used for the measurement. The
:MEASure:DEFine command has no effect on these delay measurements. The edges
specified by the :MEASure:DEFine command only define the edges used by the
:MEASure:DELay? query.
Содержание InfiniiVision 3000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 152: ...152 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Страница 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 306: ...306 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 14 EXTernal Trigger Commands...
Страница 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Страница 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Страница 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Страница 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Страница 466: ...466 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 20 MEASure Power Commands...
Страница 502: ...502 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 22 POD Commands POD1 DISP 0 THR 1 40E 00...
Страница 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Страница 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Страница 1046: ...1046 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 34 Obsolete and Discontinued Commands...
Страница 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Страница 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Страница 1086: ...1086 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 37 Synchronizing Acquisitions...
Страница 1216: ...1216 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 39 Programming Examples...
Страница 1231: ...Index Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 1231 zoomed window horizontal scale 841...
Страница 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...