29-45
Advanced Delay By Event Mode and Commands
You can set the delay mode to delay by events or time. Use Delay By
Event mode to view pulses in your waveform that occur a number of
events after a specified waveform edge. Infiniium Oscilloscopes identify
a trigger by arming on the edge you specify, counting a number of events,
then triggering on the specified edge.
Arm On
Use Arm On to set the source, level, and slope for arming the trigger
circuitry. When setting the arm level for your waveform, it is usually best
to choose a voltage value that is equal to the voltage value at the mid
point of your waveform. For example, if you have a waveform with a
minimum value of 0 (zero) volts and a maximum value of 5 volts, then
2.5 volts is the best place to set your arm level. The reason this is the
best choice is that there may be some ringing or noise at both the 0 volt
and 5 volt levels that can cause false triggers.
When you adjust the arm level control, a horizontal dashed line with a T
on the right-hand side appears showing you where the arm level is with
respect to your waveform. After a period of time the dashed line will
disappear. To redisplay the line, adjust the arm level control again, or
activate the Trigger dialog.
Delay By Event
Use Delay By Event to set the source, level, and edge to define an event.
When setting the event level for your waveform, it is usually best to
choose a voltage value that is equal to the voltage value at the mid point
of your waveform. For example, if you have a waveform with a minimum
value of 0 (zero) volts and a maximum value of 5 volts, then 2.5 volts is
the best place to set your event level. The reason this is the best choice
is that there may be some ringing or noise at both the 0 volt and 5 volt
levels that can cause false triggers.
Event
Use Event to set the number of events (edges) that must occur after the
oscilloscope is armed until it starts to look for the trigger edge.
Trigger On
Use Trigger On to set the trigger source and trigger slope required to
trigger the oscilloscope. Each source can have only one level, so if you
are arming and triggering on the same source, only one level is used.
Содержание Infiniium 8000A
Страница 1: ...Agilent Technologies Infiniium 8000A Programmer s Reference ...
Страница 2: ......
Страница 20: ...Contents 16 ...
Страница 21: ...1 Introduction to Programming ...
Страница 43: ...2 LAN and GPIB Interfaces ...
Страница 53: ...3 Message Communication and System Functions ...
Страница 58: ...3 6 ...
Страница 59: ...4 Status Reporting ...
Страница 64: ...4 6 Status Reporting Status Reporting Data Structures Figure 4 2 Status Reporting Data Structures ...
Страница 78: ...4 20 Figure 4 3 Status Reporting Decision Chart ...
Страница 79: ...5 Programming Conventions ...
Страница 84: ...5 6 Programming Conventions The Command Tree Figure 5 1 Command Tree ...
Страница 85: ...5 7 Programming Conventions The Command Tree Figure 5 2 Command Tree Continued ...
Страница 86: ...5 8 Programming Conventions The Command Tree Figure 5 3 Command Tree Continued ...
Страница 87: ...5 9 Programming Conventions The Command Tree Figure 5 4 Command Tree Continued ...
Страница 88: ...5 10 Programming Conventions The Command Tree Figure 5 5 Command Tree Continued ...
Страница 89: ...5 11 Programming Conventions The Command Tree Figure 5 6 Command Tree Continued ...
Страница 94: ...5 16 ...
Страница 95: ...6 Sample Programs ...
Страница 149: ...7 Acquire Commands ...
Страница 176: ...7 28 Acquire Commands SRATe AUTO ...
Страница 177: ...8 Bus Commands ...
Страница 187: ...9 Calibration Commands ...
Страница 195: ...10 Channel Commands ...
Страница 223: ...11 Common Commands ...
Страница 247: ...12 Digital Commands ...
Страница 254: ...12 8 ...
Страница 255: ...13 Disk Commands ...
Страница 300: ...13 46 Disk Commands STORe Obsolete ...
Страница 301: ...14 Display Commands ...
Страница 322: ...14 22 ...
Страница 323: ...15 External Trigger Commands ...
Страница 343: ...16 Function Commands ...
Страница 382: ...16 40 ...
Страница 383: ...17 Hardcopy Commands ...
Страница 391: ...18 Histogram Commands ...
Страница 403: ...19 InfiniiScan ISCan Commands ...
Страница 421: ...20 Limit Test Commands ...
Страница 429: ...21 Marker Commands ...
Страница 452: ...21 24 ...
Страница 453: ...22 Mask Test Commands ...
Страница 499: ...23 Measure Commands ...
Страница 636: ...23 138 Measure Commands VUPPer ...
Страница 637: ...24 Pod Commands ...
Страница 642: ...24 6 ...
Страница 643: ...25 Root Level Commands ...
Страница 645: ...25 3 STORe SETup STORe WAVeform TER Trigger Event Register VIEW ...
Страница 674: ...25 32 ...
Страница 675: ...26 Self Test Commands ...
Страница 679: ...27 System Commands ...
Страница 694: ...27 16 ...
Страница 695: ...28 Time Base Commands ...
Страница 708: ...28 14 ...
Страница 709: ...29 Trigger Commands ...
Страница 822: ...29 114 ...
Страница 823: ...30 Waveform Commands ...
Страница 893: ...31 Waveform Memory Commands ...
Страница 902: ...31 10 ...
Страница 903: ...32 Error Messages ...
Страница 914: ...32 12 ...