Страница 1: ...D R A F T 1 1 3 1 7 Vacuum Products Division INSTRUCTION MANUAL IDP 10 Dry Scroll Vacuum Pump Manual No X3810 90000 Revision B January 2017...

Страница 2: ...D R A F T 1 1 3 1 7 Copyright 2017 Agilent Inc IDP 10 Dry Scroll Vacuum Pump Slo Blo and Littelfuse are registered trademarks of Littelfuse Inc...

Страница 3: ...cluding representatives of Seller which are incon sistent or in conflict with the terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an officer of Seller...

Страница 4: ...IDP 10 Dry Scroll Vacuum Pump This page intentionally left blank D R A F T 1 1 3 1 7...

Страница 5: ...cal Connections 17 Start Current 18 Mechanical Connections 19 Operation 20 Cleaning the Pump 20 Startup Procedure 20 Shutdown Procedure 20 Troubleshooting 21 Maintenance 22 Kits and Service Options 22...

Страница 6: ...IDP 10 Dry Scroll Vacuum Pump This page intentionally left blank D R A F T 1 1 2 1 7...

Страница 7: ...and storing the pump the following environmental requirements should not be exceeded Preparation for Installation The pump is supplied in a special protective packing If this shows signs of damage wh...

Страница 8: ...e cable being used FCC This device complies with Part 15 of the FCC rules Oper ation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must...

Страница 9: ...hour meter which displays run time in hours Remote user control of pump speed through an analog input via the Dsub9 I O connector Automatic AC mains voltage sensing and selection for worldwide applica...

Страница 10: ...ound pins The speed varies con tinuously from full speed of 1900 RPM at 10 VDC to 1350 RPM at 6 9 VDC The pump runs at 1350 RPM for all voltages between 1 0 and 6 9 V For voltages below 1 V or without...

Страница 11: ...d for a fixed time full speed is again allowed The reduced speed setting is 1740 RPM Overtemperature Protection A thermistor is mounted to the motor housing that shuts off the power to the motor if th...

Страница 12: ...he device whether the Parent Company or a retailer to initiate the collection and disposal process after checking the contractual terms and conditions of sale For more information refer to http www ag...

Страница 13: ...erference that may cause undesired oper ation If this equipment does cause harmful interference to radio or television reception which can be determined by turn ing the equipment off and on the user i...

Страница 14: ...pompe les exigences environnementales suivantes ne doivent pas tre d pass es Pr paration pour l installation La pompe est livr e dans un emballage protecteur sp cial Si elle manifeste des signes de d...

Страница 15: ...ssez large pour int grer le c ble tant utilis FCC Cet appareil est conforme la Partie 15 des r gles de la FCC L op ration est sujette aux deux conditions sui vantes 1 ce dispositif ne provoque pas d i...

Страница 16: ...v es de la pompe Vanne d admission en option pour prot ger les sys t mes de vide de la ventilation accidentelle si la pompe s arr te l improviste Interface de l utilisateur Il y a un interrupteur coul...

Страница 17: ...ve l cran affiche 0 4 Faites glisser l interrupteur la position I ou appli quez le signal ON distant La pompe d marre apr s quelques secondes et le ventilateur s allume Si la pompe dispose de la vanne...

Страница 18: ...e en utilisant l entr e distance Entretien Le personnel responsable de l op ration et de l entretien de la pompe doit tre bien form et conscient des r gles de pr vention des accidents Mise au rebut Si...

Страница 19: ...ion Female 1 4 National Pipe Thread NW16 Gas ballast Female 1 8 National Pipe Thread 20 micron sintered plug provided Ambient operating temperature 5 C to 40 C 41 F to 104 F Storage temperature 20 C t...

Страница 20: ...IDP 10 Dry Scroll Vacuum Pump 14 D R A F T 1 1 3 1 7 Figure 4 Outline Drawing with Dimensions Figure 5 Pumping Speed Curves...

Страница 21: ...8 Main Cowling 9 Mounting Holes 4 through holes for 1 4 or M6 Hardware 10 Rear Cowling 11 KA40 Phillips PT hardware 6 12 On Off Switch 13 IO Control 14 Power Connection IEC 320 15 Hour Meter Unpacking...

Страница 22: ...the pump be sure the electrical supply is disconnected The gas ballast must be sealed when ever pumping any gas not intended to be vented to the atmosphere An exhaust silencer option is avail able for...

Страница 23: ...carry the current your product will draw An under sized cord will cause a drop in line voltage resulting in loss of power and overheating Table 6 IDP 10 Power Cord Selection Country Power Cord Specifi...

Страница 24: ...the typical IDP 10 maximum starting current and duration Table 7 Typical IDP 10 Maximum Starting Current A Frequency High Voltage Range 100 V 120 V 127 V 200 V 230 V 240 V 60 Hz 49 50 50 25 N A 38 50...

Страница 25: ...s 1 and 2 Gas ballast port 1 is used for moderate water vapor loads such as intermittently pumping out a volume exposed to normal ambient humidity Gas ballast port 2 is used for high water vapor loads...

Страница 26: ...ke sure that the pump is configured for the mains voltage to which the pump is connected Figure 9 ON OFF Power Switch 2 Switch the pump ON 3 If the pump is equipped with an optional inlet valve the va...

Страница 27: ...mperature Debris in pump Check intake screen Flush pump Disassemble pump and inspect Pump runs intermittently Motor thermal protector is cycling open and closed Allow the motor to cool Identify the ca...

Страница 28: ...nute or two to flush it out For more information see Cleaning the Pump on page 20 Tip Seal Replacement The parts and tools required to replace tip seals include Tip Seal Replacement Kit P N X3807 6700...

Страница 29: ...al onto the Orbiting Scroll item 9 3 Sequentially insert the seal from center to the outer edge of the scroll wall 4 Cut the Tip Seal about 1 8 3 mm from the groove end 5 Use the remaining Tip Seal ma...

Страница 30: ...2 Disassemble and repair as necessary The pump is now ready to return to service 1 Front Cowling Bolts M8 3 2 Exhaust Fitting NW16 3 Front Cowling Assembly 4 Frame Bolts M5 4 5 Scroll Housing 6 Main...

Страница 31: ...the core product e g inlet screens vent valves o Prior to shipment and if applicable for your product drain any oils or other liquids purge or flush all gasses and wipe off any excess residue o If ord...

Страница 32: ...15 business days Failure to do so or returning a non rebuildable part crashed will result in an invoice for the non returned non rebuildable part o Returns for credit toward the purchase of new or ref...

Страница 33: ...me Chemical name Chemical Symbol CAS Number Toxic YES NO Harmful YES NO Corrosive YES NO Reactive YES NO Flammable YES NO Explosive YES NO Radioactive YES NO Biological YES NO Oxidizing YES NO Sensiti...

Страница 34: ...IDP 10 Dry Scroll Vacuum Pump This page intentionally left blank D R A F T 1 1 2 1 7...

Страница 35: ...IDP 10 Dry Scroll Vacuum Pump This page intentionally left blank...

Страница 36: ...0574 Mexico Agilent Technologies Concepcion Beistegui No 109 Col Del Valle C P 03100 Mexico D F Tel 52 5 523 9465 Fax 52 5 523 9472 China A Agilent Technologies China Co Ltd No 3 Wang Jing Bei Lu Cha...