Agilent InfinityLab LC Series Vialsamplers User Manual
82
3
Using the Module
Install the Optional Integrated Column Compartment
Install a Column in the ICC
For more information on recommended capillary types, refer to
Tools required
p/n
Description
8710-0510
Open-end wrench 1/4 — 5/16 inch
Parts required
p/n
Description
Column (up to 30 cm length)
G7116-68003
Column Holder Lamella, 2/pk
Preparations
•
Switch off the pump
•
Switch off the sampler
•
Open the doors of the sampler
NOTE
A maximum of two columns can be installed.
1
Push the latch of the Integrated Column Compartment
to the left.
2
Open the flap door.