![Agilent Technologies G7129A Скачать руководство пользователя страница 226](http://html1.mh-extra.com/html/agilent-technologies/g7129a/g7129a_user-manual_2867753226.webp)
Agilent InfinityLab LC Series Vialsamplers User Manual
226
8
Maintenance
Overview of Maintenance
Overview of Maintenance
It is necessary to perform periodic inspection of this instrument to ensure its safe
use. It is possible to have these periodic inspections performed by Agilent service
representatives on a contractual basis. For information regarding the
maintenance inspection contract, contact your Agilent representative.
The following pages describe the maintenance (simple repairs) of the module
that can be carried out without opening the main cover.
Table 21
Overview of maintenance
Procedure
Typical interval (minimum)
Change needle/needle seat
30000 needle into seat movements
Change peristaltic pump cartridge
3000 h on time
Change rotor seal
30000 injections
Change metering seal
30000 injections