Agilent InfinityLab LC Series Vialsamplers User Manual
132
3
Using the Module
Transporting the Sampler
Transporting the Sampler
Prepare a Sampler Without Cooler/Thermostat for
Transportation
1
Remove all vials from the vial drawers.
2
Move the transport assembly to the park position using Instant Pilot or Lab
Advisor, see
3
Turn off the sampler.
4
Install the Transport Protection Foam, see
“Install the Transport Protection
When
The module needs to be transported or relocated.
WARNING
Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy
objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
CAUTION
Unsecured transportation
Mechanical damage
Secure the transport assembly before transporting the sampler.