Agilent InfinityLab LC Series Vialsamplers User Manual
125
3
Using the Module
Using the Optional Sample Cooler/Sample Thermostat
Control
With the Sample Cooler/Sample Thermostat installed, the
Control
dialog box of
the hosting Infinity II sampler will include the following
cooler/thermostat-specific control options:
• At Power On
:
•
Turn On Thermostat
: The cooler/thermostat turns on automatically upon
powering on the sampler.
• Thermostat
:
•
On
: The cooler/thermostat turns on and the system starts to regulate the
temperature inside the sample space towards the setpoint.
•
Off
: The cooler/thermostat turns off.
• Enable Analysis
•
With any temperature
: The analysis starts regardless of the actual
temperature inside the sampler.
•
Temperature /- 2 °C
: The analysis starts only when the actual
temperature is within the ± 2 °C range of the setpoint temperature.
NOTE
For the Sample Cooler, the set temperature must be at least 5 °C below ambient
for proper temperature control.
NOTE
The
Enable Analysis
control setting is available since LC & CE drivers A.02.19.
NOTE
The
Temperature /- 2 °C
option is only available for the Sample
Thermostat.