Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
213
Maintenance
9
Exchanging a Flow Cell
Exchanging a Flow Cell
For bio-inert modules use bio-inert parts only!
When
If an application needs a different type of flow cell or the flow cell is defective (leaky).
Tools required
p/n
Description
Wrench, 1/4 inch
for capillary connections
OR
5043-0915
Fitting mounting tool
Parts required
p/n
Description
G1321-60005
Flow cell, 8 µL, 20 bar (pH 1 – 9.5)
G1321-60015
Flow cell, 4 µL, 20 bar (pH 1 – 9.5)
G5615-60005
Bio-inert flow cell, 8 µL, 20 bar (pH 1–12) includes Capillary Kit Flow Cells BIO
(p/n G5615-68755)
G1321-60007
FLD Cuvette Kit, 8 µL, 20 bar
Preparations
Turn off the flow.
C A U T I O N
Sample degradation and contamination of the instrument
Metal parts in the flow path can interact with the bio-molecules in the sample leading
to sample degradation and contamination.
➔
For bio-inert applications, always use dedicated bio-inert parts, which can be
identified by the bio-inert symbol or other markers described in this manual.
➔
Do not mix bio-inert and non-inert modules or parts in a bio-inert system.
Содержание G7121A
Страница 86: ...86 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 4 Preparing the Module Solvent Information ...
Страница 230: ...230 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 10 Parts for Maintenance and Repair Kits ...
Страница 264: ...264 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 12 Hardware Information Early Maintenance Feedback ...
Страница 281: ...Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 281 Index ...