![Agilent Technologies G7121A Скачать руководство пользователя страница 191](http://html.mh-extra.com/html/agilent-technologies/g7121a/g7121a_user-manual_2867752191.webp)
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
191
Test Functions and Calibration
8
Excitation and Emission Grating Resistance History
Excitation and Emission Grating Resistance History
This test runs automatically when the instrument is turned on (not accessible
as an external test).
It provides the resistance history of the excitation and the emission grating
drives. The number of revolutions after switching off the drives is a measure
of friction. The history may show an increasing friction of the drive(s) over a
length of time.
The history data contains the data/time information and the number of turns.
The data/plot can be retrieved via the diagnostics and provides turn data over
a length of time.
Figure 70
Resistance History (Agilent Lab Advisor under
Module Info
)