Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
187
Test Functions and Calibration
8
Wavelength Calibration Procedure
Wavelength Calibration Procedure
1
Preparation of the Glycogen Calibration Sample.
a
To prepare 10 ml of the calibration solution you have to use 10 mg of the
Glycogen sample (a tolerance of ±20% is not critical).
b
Fill the prepared amount into a suitable bottle/vial.
c
Fill 10 ml of distilled water into the vial and shake.
d
Wait 5 minutes and shake again. After 10 minutes the solution is ready.
When
If application requires, or see
Tools required
Description
Laboratory balance
Parts required
p/n
Description
G7121-68001
FLD Calibration Kit
contains parts below
5063-6597
Calibration Sample, Glycogen
9301-1446
Syringe
9301-0407
Needle
5190-5111
Syringe filter, 0.45 µm, 100/pk
0100-1516
Fitting male PEEK, 2/pk
N O T E
New instruments are shipped with a flow cell installed and have a factory wavelength
calibration.
If a flow cell is used other then the installed/shipped flow cell:
•
verify that the installed flow cell is fixed very tight,
•
perform a wavelength calibration using Glycogen (part of FLD Calibration Kit
(G7121-68001)).
Содержание G7121A
Страница 86: ...86 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 4 Preparing the Module Solvent Information ...
Страница 230: ...230 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 10 Parts for Maintenance and Repair Kits ...
Страница 264: ...264 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 12 Hardware Information Early Maintenance Feedback ...
Страница 281: ...Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 281 Index ...