18
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
1
Introduction to the Fluorescence Detector
Raman Effect
Raman Effect
The Raman effect arises when the incident light excites molecules in the
sample which subsequently scatter the light. While most of this scattered light
is at the same wavelength as the incident light, some is scattered at a different
wavelength. This inelastically scattered light is called Raman scatter. It results
from the molecule changing it's molecular motions.
Figure 6
Raman
The energy difference between the incident light (E
i
) and the Raman scattered
light (E
s
) is equal to the energy involved in changing the molecule's vibrational
state (i.e. getting the molecule to vibrate, E
v
). This energy difference is called
the Raman shift.
E
v
= E
i
- E
s
Several different Raman shifted signals will often be observed; each being
associated with different vibrational or rotational motions of molecules in the
sample. The particular molecule and its environment will determine what
Raman signals will be observed (if any).
A plot of Raman intensity versus Raman shift is a Raman spectrum.
6DPSOH
,QFLGHQW/LJKW
5DOHLJK6FDWWHU
VDPHZDYHOHQJWK
DVLQFLGHQWOLJKW
5DPDQ6FDWWHU
QHZZDYHOHQJWK
6FDWWHUHG/LJKW
Содержание G7121A
Страница 86: ...86 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 4 Preparing the Module Solvent Information ...
Страница 230: ...230 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 10 Parts for Maintenance and Repair Kits ...
Страница 264: ...264 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 12 Hardware Information Early Maintenance Feedback ...
Страница 281: ...Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 281 Index ...